Power frequency magnetic field immunity test system PMF-801C Return
Product overview
PMF-801C Series power-frequency magnetic field immunity testing system is developed and manufactured by LIONCEL, which meets the requirements of the latest IEC 61000-4-8 standards and GB/T17626.8-2006 standards. It can fulfill all the test requirements specified in the standards.

Feature

1.Modular design, can extended for other test such as pulse magnetic field immunity test and damped oscillation magnetic field immunity test. 

2. Color touchscreen control, built-in IEC standard level; 

3. RS485 host computer port, realize remote control.

4.  Advanced coil technology and professional material selection.

5.  Distortion rate of magnetic field is less than 8%.

 

Meet the standards:

IEC61000-4-8 EN61000-4-8 GB/T 17626.8

application area:
technical parameter:

Magnetic field intensities

1A/m 100A/m continually adjustable continuous mode

100 A/m1000A/m continuous adjustable

short time operation5s

Total distortion rate of output current

8%

Duration of magnetic field

1s999minlong-term

1s5sshort-term

Interval of magnetic field

1s99min

Magnetic direction

Horizontal, vertical, 360 °rotation, and can move up and down (0-1.8 meters), support proximity method and immersion method.

Magnetic field frequency

50Hz or 60Hz

parts: