Technical characteristics
■ automatic coupling decoupling network, free switching common differential mode test, one key test can be realized by setting common differential mode ratio;
■ external network, can cooperate with any brand lightning surge / ringing wave generator to complete automatic lightning surge test;
■ modular architecture, easy to maintain and upgrade.
IEC61000-4-5 ed.3 C61000-4-12
Single phase coupling/decoupling network |
|
Parameter |
CDN-511P |
Coupling surge wave |
Open circuit voltage wave Raise time:1.2μs±30% Width:50μs±20%
Short circuit currency wave Raise time:8μs±20% Width:20μs±20% |
Max input voltage |
6.6kV |
Max input currency |
3.3kA |
Phase synchronization |
Asynchronous/synchronous 0 to 359° |
Power supply |
Open circuit voltage wave Raise time:1.2μs Width:50μs±20%
Short circuit currency wave Raise time:8μs Width:20μs±20% |
EUT capacity |
AC220V
|
Size |
19''/4U |
Weight |
20kg |