Products
Products
FFT3010_3030EMI measurement receiver Return
Product overview
Full FFT digital electromagnetic interference test receiver can measure 9KHz-300MHz conducted electromagnetic interference

FFT 3010 & 3030 include preselector and preamplifier and advanced EMC testing software. The PC integrated architecture of Windows 10 embedded operating system, FFT 3010 & 3030 EMI receivers can be controlled by advanced software to achieve EMI testing. The built-in preselector provides excellent dynamic range and accurate electromagnetic radiation interference measurement. The frequency range covers 9 KHz-300 MHz. The remote control can also be performed by an external PC.

  The built-in pre-selected amplifier of FFT3010 & 3030 has excellent dynamic measurement range characteristics and provides extremely accurate electromagnetic compatibility test measurement. It meets the requirements of international electromagnetic interference, European and product standards, and compares international standards through measurement results The limit value and the corresponding detector measurement value are used to complete the test.

main feature:

  FFT (Fast Fourier Transform) scan mode

  Peak, quasi-peak, CISPR average

  RMS (effective value) and CISPR RMS digital detector

  Automatic insertion attenuation during measurement scan to avoid receiver saturation

  The analysis function has an accurate digital overload detector to avoid the influence of saturation

  Modified pulse weighting according to CISPR16-1-1 (from PRF 1Hz)

  Fast test, fast detection of critical frequency (dwell time adjusted to 1ms)

  High sensitivity

  Strong signal anti-interference

  Low measurement uncertainty

  High test speed

  Test cable insertion loss, attenuator, amplifier loss, coupling network correction value, GTEM correction, antenna factor correction

  Integrated signal generator

  External reference frequency: 10MHz

  Digital demodulation option with AM/FM/WBFM



Meet the standards:
application area:
technical parameter:



FFT3010

FFT3030

 

frequency

 

Frequency Range

9kHz-30MHz

9kHz-300MHz

 

Frequency setting

1Hz

1Hz

 

Internal reference frequency

 

Aging per Year

2 x 10-6

2 x 10-6

 

Temperature drift

15 x 10-5 (+10 °C to +40°C)

15 x 10-5 (+10 °C to +40 °C)

 

External reference frequency

10MHz

10MHz

 

Measuring time (manual mode)

1ms to 5s

1ms to 5s

 

Resolution

1ms

1ms

 

Measuring time (scan mode)

1ms to 5s

1ms to 5s

 

Resolution

1ms

1ms