Current injection probe calibration fixture FCC-BCICF-6-150 Return
Product overview
FCC-BCICF-6-150 is designed to meet the standard IEC 61000-4-6 Ed 4.0 section 6.2.3.1–6.2.3.2, which stipulates that "...when the current injection clamp is inserted, the test fixture transmission loss shall not increase More than 1.6dB". This 1.6dB requirement is verified by subtracting the insertion loss when the calibration fixture is not in place from the insertion loss when the calibration fixture is in place.

 FCC-BCICF-6-150 is designed to meet the standard IEC 61000-4-6 Ed 4.0 section 6.2.3.1–6.2.3.2, which stipulates that "...when the current injection clamp is inserted, the test fixture transmission loss shall not increase More than 1.6dB". This 1.6dB requirement is verified by subtracting the insertion loss when the calibration fixture is not in place from the insertion loss when the calibration fixture is in place.

Frequency range: 150kHz-230MHz

  Insertion loss: <1.6dB

  Length: 432mm

  Width: 229mm

  Height: 80mm

  Weight: 2.2kg

  Optional accessories: Passive impedance adapter FCC-801-150-50-BCI


Meet the standards:
application area:
technical parameter:

Typical insertion loss

parts:
Video playback: