IGUU2916 EMI calibration pulse and sine wave generator (9k-1G) Return
Product overview
IGUU 2916 is used by various EMI receiver manufacturers and measurement organizations, and is the most widely used calibration pulse source in the world.

    When calibrating an electromagnetic interference measurement receiver, especially when calibrating the impulse response characteristics of the receiver according to the CISPR-16 standard, a pulse signal generator must be used. IGUU 2916 is used by various EMI receiver manufacturers and measurement organizations, and is the most widely used calibration pulse source in the world.

main feature:

      When calibrating an electromagnetic interference measurement receiver, especially when calibrating the impulse response characteristics of the receiver according to the CISPR-16 standard, a pulse signal generator must be used. IGUU 2916 is used by various EMI receiver manufacturers and measurement organizations, and is the most widely used calibration pulse source in the world.

       IGUU 2916 consists of three signal sources: a main pulse generator, an auxiliary pulse generator and a sine wave generator. Since the calibration pulse generator itself has a sinusoidal signal generator, there is no need to connect an external signal source when calibrating the pulse response characteristics of the EMI receiver. Through the dedicated interface, test software and coaxial switch, the impulse response characteristic calibration of the electromagnetic interference receiver can be automatically completed. The control software is based on WINDOWS operating system, with friendly interface and easy operation.

       The optional coaxial switch unit KU 9616, which can be controlled by IGUU 2916, connects one of the four outputs of IGUU 2916 to the equipment under test (EUT). There is no need to switch the coaxial connector during the test, which is convenient and time-saving.

       Through the GPIB interface, the computer can fully control all the functions of IGUU 2916. If KU 9616 is selected, fully automatic testing can be realized.


Meet the standards:
application area:
technical parameter:

1. Main generator:

Pulse frequency range:

0.1 Hz-20 Hz, 1-200 Hz, 10 Hz-2 kHz, 100 Hz -20 kHz, in steps of 0.1 Hz, 1 Hz, 10 Hz, 100 Hz

Frequency accuracy: ±2×10 -4  Hz

Pulse shape: almost square wave

Pulse width: Band A 270ns; Band B 6ns; Band C/D 0.3ns

Pulse polarity: positive /negative (switchable)

Pulse amplitude: (CISPR weighted QP):

Band A / B 10-70dB in steps of 1dB

Band C/D 0-60 dB in steps of 1dB

Maximum output voltage (across 50Ω load): 80V~110V

Setting of Pulse Area at 60dB attenuator:

Corresponds to the CISPR 16 standard QP detector, which is equal to 1mV rms at 50 ohms .

Frequency flatness:

Typical value: ±0 to -0.5dB at the high end frequency of band A/B

Band C: Typical value: ±0.5dB (30MHz-100MHz) (100MHz is the reference frequency for level calibration)

The level is between 200MHz and 800MHz, with the frequency increasing from +0.5dB to +1dB, and when the frequency is 800MHz-1000MHz, it decreases to ±0 to -1dB.

Level accuracy:

Band A/B: ±0.5 dB at 50 kHz, 1 MHz

Band C/D: ±0.8 dB at 100 MHz

 

2. Auxiliary generator:

Pulse frequency range:

0.1 Hz-20 Hz, 1-200 Hz, 10 Hz-2 kHz, 100 Hz -20 kHz, in steps of 0.1 Hz, 1 Hz, 10 Hz, 100 Hz

Pulse frequency accuracy: ±2×10 -4  Hz

Pulse width: Band A 70ns; Band B 5ns; Band C/D 1ns

Pulse amplitude: ( CISPR weighted QP):

Band A/BC/D: 40dB fixed level

Frequency flatness:

Band A/B: Same as main generator

Band C (D): typical value ±0.75 dB, 30MHz-300MHz; -3 dB >500 MHz

 

3. Dimensions: W x H x D

447mm x 180mm x 460mm