Product overview
LEG-433C series pulse group generator is Lingshi
  • Universal power supply range design to meet the needs of different countries and regions;
  • The instrument is equipped with high voltage power supply overvoltage protection, over-temperature protection, and sound/light prompts;
  • Rich and complete communication interface design, standard configuration of RS485 serial port, LAN port, USB port, wireless WiFi;
  • Emergency stop switch: quickly cut off the internal high-voltage power supply of the instrument or the EUT power supply circuit in an emergency;
  • Safety lock: The device can be operated normally when the safety lock is inserted, and the device is locked when the safety lock is removed;
  • External synchronization: used to connect an external manual coupling and decoupling network to achieve EUT power supply phase synchronization function;
  • Manual trigger / automatic trigger: manually control the instrument pulse output through the virtual buttons on the touch screen, output a series of pulses for each point, and automatically trigger the pulse output to be controlled by the set program;
  • External trigger: The pulse output can be controlled by connecting an external trigger switch (normally open).
  • EUT_FAIL failure detection, can provide three monitoring inputs.


Meet the standards:
IEC 61000-4-4 IEC 61000-4-4 Ed.3/Ed.2 EN 61000-4-4 GB/T 17626.4
application area:
Industrial equipment, power meters, vehicle electronics, medical equipment, lighting fixtures, communication transmission, audio and video equipment, low-voltage electrical appliances, electronic components, power tools, information technology equipment, rail transit and aerospace electrical appliances, instruments and meters
technical parameter:

Model

LEG-433C-05AC

Interference Type

Electrical fast transient burst immunity

power supply
AC 85 V ~ 264 V 50/60 Hz 120 W
Fuse capacity: 2 A/250 V
Dimensions
450 mm×545 mm×195 mm
Signal Source
Pulse voltage range
100 Vp ~ 5100 Vp±10 %
Pulse rise time
50 Ω&1 kΩ load: rise time 3.5 ns to 6.5 ns
Pulse duration
50 Ω load: 35 ns to 65 ns
1 kΩ load: 35 ns to 150 ns
Output Impedance

50 Ω/coaxial output

polarity
Positive, negative, positive and negative automatic alternation
Repetition frequency
0.1 ~ 1000 kHz
Number of single group pulses
1 to 1000
Group Period
10 to 9999 ms

Built-in coupling/decoupling network

aisle
Three-phase five-wire
Host computer interface
LAN/Wireless WiFi
Test sample capacity

AC 690V&32A MAX/DC 1000V&32A MAX

Pulse rise time
50 Ω load: rise time 4 ns to 7 ns
Pulse duration
50 Ω load: 30 ns to 60 ns
parts:
Video playback:

Copyright:Shanghai Lioncel Electromagnetic Technology Co., Ltd (2020)